Description
What it is
A 10 A power supply with a backup battery, in a lockable metal case. We recommend it for any nurse call system with a WAC-07 Wireless Annunciator Controller and annunciator displays. One supply runs up to 4 single-sided or 3 double-sided displays, plus the controller, so you don’t need a plug pack for each display.
What’s included
- 1 ร 10 A power supply in a lockable metal case
- 1 ร 12 V 7 Ah sealed lead acid backup battery, fitted
- Fused output for annunciator displays, fitted
- 2 ร keys
- 1 ร spare 10 A blade fuse
Features
- Display output: 10 A for the displays, protected by an inline blade fuse and wired to lever terminal blocks.
- 18 fused outputs for the WAC-07, RS485 splitter and data logger. Each output has a resettable fuse and an LED, so a fault on one device doesn’t take down the rest.
- Battery backup: when mains power fails, the battery takes over. It recharges once power is back.
- Lights on the door show Power, DC 12 V and Charge at a glance.
- Two switches: one for mains power, one for the DC output.
- Protection against short circuit, high voltage, over current and overheating.
- Lockable metal case with room to tidy the cables.
- BT STUP button to start the supply from the battery alone.
Connecting it
- Turn the power switch off.
- Connect the red and black leads to the battery: red to positive, black to negative.
- Connect the WAC-07 and other central equipment to the 550 mA outputs, one device per output. The top row is positive and the lower row is negative.
- Connect the annunciator displays to the display output terminal blocks: red is positive, black is negative.
- Plug the supply into a 240 V AC power point.
- Turn the power switch on. Each output LED and the battery charge LED should light.
- Close the cover and lock it with the key.
As the battery charges, the charge LED on the front panel gets brighter.
Specifications
| Part number | N-POWER-SUPPLY-10AMP |
| Model | Securview PS12VDC10A18CHB |
| Input | 100 to 240 V AC, auto switching |
| Output voltage | 13.8 V DC |
| Total output current | 10 A |
| Displays | Up to 4 single-sided or 3 double-sided |
| Display output | Inline blade fuse, lever terminal blocks |
| Equipment outputs | 18, each 0.55 A with a resettable fuse and LED |
| Backup battery | 12 V 7 Ah sealed lead acid (fitted) |
| Protection | Short circuit, high voltage, over current, thermal cut-off |
| Case | Metal, with key lock |
| Dimensions | 265 ร 205 ร 90 mm |
| Use | Indoor only |
| Warranty | 12 months |
Works with
Frequently asked questions
Do I need this, or will a plug pack do?
One display can run from the plug pack that comes with the display kit. Once you add a second display, it needs power too, and this supply covers up to 4 single-sided or 3 double-sided displays with battery backup.
How long will it run in a power failure?
That depends on how many displays you have. These are estimates, based on each device’s maximum current draw and the 7 Ah battery:
| WAC-07 and 1 single-sided display | about 5 hours |
| WAC-07 and 2 single-sided displays | about 3 hours |
| WAC-07 and 4 single-sided displays | about 1.5 hours |
| WAC-07 and 2 double-sided displays | about 1.5 hours |
| WAC-07 and 3 double-sided displays | about 1 hour |
Real times vary with the battery’s condition and how many calls are on the displays.
How much does each device draw?
As a guide: a single-sided display about 1.1 A, a double-sided display about 2 A, the WAC-07 controller 200 mA, the RS485 splitter 200 mA and the data logger 360 mA. Send us your layout and we’ll check it for you.
Can I run display power in the data cable?
Yes, for single-sided and double-sided displays up to 80 to 100 m away, in the CAT5/6 cable. Further than that, voltage drop is too high, so the display needs its own power. You can also run power in separate figure 8 or security cable, at least 14/0.2 mm stranded. Any display powered separately must share a common ground with the rest of the system.
Why does a 12 V call point show “-LB-” on its first call?
Call points don’t usually run from this supply. If you do power a call point from 12 V, it shows “-LB-” (low battery) on its first transmission after it’s powered up. You can ignore it: it won’t appear on later calls.
How do I reset the system if a display freezes?
Open the case and flick the 12VDC Output ON/OFF switch off, then back on. Switching off at the power point won’t reset it, because the battery keeps the system running.
What about a small system with no displays?
If your call points send straight to pagers with no WAC-07, the PowerShield DC Mini UPS can back up each 12 V call point instead.
Can it be used outdoors?
No, it’s for indoor use only.
Is it GST-free?
It’s GST-free for eligible medical and disability use. Contact us for a quote.